Inductively coupled plasma mass spectrometer (ICPMS)

time:2021-06-01Hits:278设置


Introduction: The inductively coupled plasma mass spectrometer is the instrument of measuring the trace content of elements and ratio of isotopes. The advantage of this instrument is that it has low detection limit (ng/ml or lower), the base effect is small,simple line, and could determine many elements at the same time. On the other hand, dynamic linear range is wide and could determine the isotope ratios rapidly.

Instrument name

Inductively coupled plasma mass spectrometer (ICPMS)

Model

 M90

Manufacturer  

analytikjena

Secifications

1. Sensitivity High sensitivity mode:

Low mass number Be (9) > 100 M cps/ppm

Medium mass number In(115)> 1000M cps/ppm

High mass number Th(232) > 500 M cps/ppm

2. Oxide ion ration

CeO+ / Ce+  2 %Normal sensitivity mode

CeO+ / Ce+  3 %High sensitivity mode

3. Bivalent ion rationBa++ / Ba+  3 %

4. background5amu)< 2 c/s

5. Short-time precision 20 min)< 3 %

6. Long-time stability4 hours 4 %

7. abundance sensitivity low-end 5*10-7, high-end 1*10-7

8. accuracy of isotope ratio 0.1 % (107Ag/109Ag)

9. element detection limitng/L 9Be0.5; 115In0.1; 232Th0.05

Applications and Uses

ICPMS is frequently used to analyze various elements.

Accessories

None

Contact

Xiaodie He

Tel.

65884616

E-mail

xdhe@suda.edu.cn

                                                                                                                                                        Edited by Xiao-Die He

返回原图
/